| MODEL | 7410 | 7420 | 7430 | 7440 |
| AC WITHSTAND VOLTAGE |
| Output Rating | 5KVAC/30mA | 5KVAC/30mA | 5KVAC/40mA | 5KVAC/40mA |
| Range | Resolution | Accuracy |
| Output Voltage,ACV | 0-5000 | 0.01 | ±( 2% of setting +5V ) |
| Output Frequency | 50Hz/60Hz±100ppm,User Selection |
| Output Waveform | Sine Wave,THD<2%( Resistion Load ),Crest Factor = 1.3-1.5 |
| Output Regulation | ±( 1% of output +5V ) |
| SETTINGS |
| Hi-Limit AC Current,mA | 0-30.00 | 0-30.00 | 0-40.00 | 0-40.00 |
| Resolution:0.001,Accuracy:±( 2% of setting +3counts ) |
| Lo-Limit AC Current,mA | 0-9.999 |
| Resolution:0.01,Accuracy:±( 2% of setting +3counts ) |
| Ramp Time,second | 0-999.9 | 0.1 | ±( 0.1% + 0.05sec ) |
| Dwell Time,second | 0,0.3-999.9 | 0.1 |
| ( 0=continuous ) |
| Arc Detection | 1-9 ranges ( 9 is the most sensitivity ) |
| AC Current Offset,mA | 0-2.000 or Auto Set |
| DC WITHSTAND VOLTAGE |
| Output Rating | 6KVDC/9999uA |
| Output Voltage,DCV | 0-6000 | 1 | ±( 2% of setting +5V ) |
| Output Ripple | < 5% ( 6KV/9999uA at Resistive Load ) |
| SETTINGS |
| Hi-Limit DC Current,uA | 0-9999 | 1 | ±( 2% of setting +3counts ) |
| Lo-Limit DC Current,uA | 0-999.9 | 0.1 | ±( 2% of setting +3counts ) |
| Ramp Time,second | 0.1-999.9 | 0.1 | ±( 0.1% + 0.05sec ) |
| Dwell Time,second | 0,0.4-999.9 | 0.1 |
| ( 0=continuous ) |
| Arc Detection | 1-9 ranges ( 9 is the most sensitivity ) |
| DC Current Offset,uA | 0-200,0 or Auto Set |
| Ramp High Current | ON/OFF,User Selection |
| Charge-Low Current,uA | 0-350,0 or Auto Set |
| Discharge Time | ≦200msec |
| INSULATION RESISTANCE |
| Output Rating | 1KVDC/9999MW |
| Output Voltage,VDC | 100-1000 | 1 | ±( 2% of setting +3V ) |
| Hi-Limit Resistance,MW | 0.1-9999(0=OFF ) | 1 | ±(2% of setting +1counts) at 500-1000VDC,1-1000MW ±(8% of setting +1counts) at 500-1000VDC,1000-9999MW ±(8% of setting +1counts) at 100-500VDC,1-1000MW |
| Lo-Limit Resistance,MW | 1-9999 | 1 |
| Delay Time,second | 0,0.5-999.9 | 0.1 | ±( 0.1% + 0.05sec ) |
| ( 0=continuous ) |
| Charge-Low Current,uA | 0-3.500 or Auto Set |
| GROOUND BOND(7440 ONLY) |
| Output Rating | 30A/600mW ,8V |
| Output AC Current | 3.00-30.00 | 0.01 | ±( 2% of setting +3counts ) |
| Output AC Voltage | 3.00-8.00 | 0.01 | ±( 2% of setting +3counts )at Open circuit |
| Output Frequency | 50Hz/60Hz±100ppm,User Selection |
| SETTINGS |
| Hi-Limit Resistance,mW | 0-150( 3-30A ) | 1 | ±( 2% of setting +2counts ) |
| Lo-Limit Resistance,mW | 0-600( 3-10A ) |
| Lead Resistance,Offset mW | 0-200 | 1 | ±( 2% of setting +2counts ) |
| Dwell Time,second | 0,0.5-999.9 sec ( 0=continuous ) | 0.1 | ±( 0.1% + 0.05sec ) |
| GENERAL |
| Input Voltage AC | 115/230VAC±15%,50/60Hz±5%,Fuse 5A Slow for7410/7412 model,6.3A Slow for7420/7430/7440 model |
| PLC Remote Control | Input: Test,Reset,( Memory1.2.3 |
| Output: Pass,Fail,Processing |
| ※Ordering Information※--------------------------------------------------------------------------------- |
| 7410 Dielectric Analyzer (ACW,DCW,IR) |
| 7420 Dielectric Analyzer with Build-in 8CH Scanner(ACW,DCW,IR) |
| 7430 Dielectric Analyzer(ACW,DCW,IR) |
| 7440 Electrical Safety Compliance Analyzer(ACW,DCW,IR,GB) |