检测离子式静电控制系统性能的简单、之方法;
合D协会No.3标准;
时间及电压分开显示,用于衰退率分析;
双档位:0~&plun;5kV;
可证实感应电荷及人体电荷之建立。
放电测试电路的电容量: 20&plun;2pF
充电板自行放电率:5分钟,设定电压的10%以下
体积:15.2*15.2c:m
输出: 6000V
面板充电电压: &plun;1000V及&plun;5000V
测试范围:0 to &plun;7500V,0 to &plun;2000V开关可调;
:高于2%;
偏移值:10分钟数值稳定后,每小时的偏移值低于0.4%;
反应速度:从10%到90%满刻度变化需不到100ms时间;
输出:将输入分成1000份,电容阻为100;
电源:220V/50HZ。
Make sure antistatic devic are working
Wrist straps, work surfac, flooring materials —
static decay measurement is a sure indication that
they’re doing their job. The plate has a connector to
permit measurement of charge buildup on operators.
Easy to use
To operate in discharge mode:
1. Locate the removable plate/probe where
ionization is to be checked, main instrument
where convenient.
2. Turn the instrument on and select tt parameters
(mode, polarity, voltage range, and upper and lower
voltage limits across which the decay rate is to be
measured.
3. Prs ZERO to ground the plate, then prs
CHARGE to bring the plate to the prelected
maximum voltage.
4. Prs DECAY, as the ionization system
neutraliz the plate charge, the instrument will
continuously measure the decaying voltage.
Elapsed time (in seconds) is displayed on the front
panel, along with plate voltage. The speed at which
the voltage drops is proportional to ionization
system performance.
Measurement in the peak mode is just as simple.