Bios Defender 530 Calibrator -- Coming January 2010
The Bios Defender 530, with the ability to take both volumetric and standardized readings, is the latest addition to the popular Defender 510 and 520 line of Bios DryCal primary flow calibrators. Automatic correion of readings for temperature and pressure, rather than merely recording of ambient conditions, is essential for many applications.
The Defender 530, like the other instruments in the line, provides users with a true primary flow calibrator for industrial hygiene, environmental and laboratory applications. The design offers low cost, light weight and ultra-compa size, incorporating the highly-regarded DryCal technology. It is all enough to fit easily into a slim briefcase for convenient transport.
Like all Bios calibrators, the Defender 530 is easy to use. Turn it on. Conne it to a sampling pump. Start taking readings. It's as simple as that. Calibrations within a minute with no messy soap bubble solutions.
The Defender 510, 520 and 530 can take hands-free continuous readings while quickly and accurately setting the flow rate on air sampling pumps or other instruments. The Defender 510/520/530 calibrators conne to any PC and pre- and post-calibration data can be downloaded with the included Integrator 110 with the Optimizer Colle Light software.
| ||||
Size | 5.5"H x 6"W x 3"D 140 mm x 150 mm x 75 mm | |||
Weight | 29 oz. / 820 g | |||
Flow Ranges | | |||
DryCal Defender Low (L) Flow: | 5 to 500 ccm | |||
DryCal Defender Medium (M) Flow: | 50 - 5,000 ccm | |||
DryCal Defender High (H) Flow: | 300 - 30,000 ccm | |||
Accuracy | | |||
Volumetric | ±1.0% | |||
Standardized | ±1.2% | |||
User-Seleable Measurement Units | ||||
Volumetric Flow | mL/min, L/min, cc/min, cf/min | |||
Standarized Flow | L/min, sL/min, scc/min, scf/min | |||
Pressure | mmHg, PSI, kPa | |||
Temperature | oC, oF | |||
Readings | ||||
Approximate Time per Reading | 1-15 seconds | |||
Reading Styles | Single (manual), Continuous or Burst, with averaging funion user-seleable from 1 to 100 measurements | |||
Gas Compatibility | Use with non-corrosive, non-condensing, non-combustible gases, less than 70% humidity | |||
Flow Modes | Suion or Pressure | |||
Measurement Cell Style | Integrated | |||
Temperature & Pressure Sensors | In the flow stream Press: 3.5 mmHg (tical), 7.0 mm (max) Temp: 0.8 °C (tical) 1.3 °C (max) | |||
Battery System | ||||
AC Adapter/Charger | 12 DC, >250 ma, 2.5 mm, center positive | |||
Battery | 6V rechargeable, sealed lead-acid, 6-8 hours tical operation | |||
Battery Operational Time | (5 cycles/min), 3 hours backlight on, 8 hours backlight off | |||
Pressure & Suion Fittings | ?” ID Hose Barb | |||
Display | Backlit graphical LCD | |||
Storage Temperature | 0-70 °C | |||
Ambient Humidity | 0-70% - non-condensing | |||
Operating Pressure | (Absolute) 15 PSI | |||
Warranty | Instrument: 1 year, Battery: 6 months | |||
Data Port | Serial RS-232 (for use with Optimizer Colle software) | |||
Data Cable | 1 meter (Defender Data Port to PC Serial Port) | |||
Proteive Case | Soft side or Pelican cases available | |||
Laboratory Accreditation | Backed by ISO 17025 accreditation by NVLAP of NIST | |||
Instrument Certification | RoHS and CE Compliant |