- 品牌/商标:THERMO
- 新旧程度:全新
防爆安全温度计(THERMO) 型号:TP9
库号:M214985
测温范围:-40 to +400°F
-40 to +204°C
误差:
±0.2°F from 32 to 200°F
±0.5°F from 200 to 400°F
±0.1°C from 0 to 100°C
±0.3°C from 100 to
200°C
2节AA碱性电池寿命约200小时
外壳材料:
不锈钢
探针
304不锈钢
尺寸:
10“×4.25”x
6.4“W
The TP9 employs the proven RTD design that has
been used in the TP7 and TP8 for many years. A sealed industrial quality overlay
provides a user interface that is easy to use with gloves. A stainless steel
enclosure protects the circuit board and large LCD from penetration by impact,
water and reactive liquids.
The new TP9 circuit board is an evolutionary step
up from our highly accurate, reliable and successful TL1 laboratory thermometer.
The Power Button’s primary function powers the instrument for intervals of about
20 minutes since the last button was accessed. The Power Button can also be used
to conserve power and clear the memory, or to make adjustment while in
calibration mode. A simple menu operation is displayed by holding the Function
Button, and alternately functions to allow adjustments in the calibration mode.
Arrows on the left side of the display show the direction of the temperature
reading and whether stability has been reached. At the user’s discretion
stabilized temperatures can be logged at numerous liquid levels for a running
average and later displayed for the user’s documenting purposes. But this
feature never interferes with simply getting an accurate temperature
reading.
To endure the environment and be intrinsically safe, the TP 9 is
manufactured of materials, which are both immune to petrochemicals and are
non-sparking. The enclosure is made of stainless steel. The probe assembly is
constructed using non-stick cable and stainless steel sensor
components.
Specifications
Maximum Dimensions:
10”L x 4.25”H x
6.4” W
Probe
304 Stainless Steel, Sealant, Aramid Fiber
Reinforced, FEP or PFA
Cable Jacket, Coaxial
Construction
Enclosure Material:
Stainless Steel
Acetal
(Delrin) Probe Holder
Polyvinylchloride Faceplate
Stainless Steel
Fasteners
Batteries:
2 AA Alkaline; Battery Life of approximately 200
hours
Battery manufacturer’s battery operating temperatature range -4 to
130°F, -20 to 54°C
Note: Battery may not provide adequate power if ambient
temperature is extremely low or high.
Temperature:
Resolution
0.01
Range: -40 to +400°F
-40 to +204°C
Calibrated Accuracy:
±0.2°F
from 32 to 200°F
±0.5°F from 200 to 400°F
±0.1°C from 0 to 100°C
±0.3°C from 100 to 200°C
4 Point NIST Traceable Report of
Test
Long-term drift not to exceed 0.05%/year
Meets API
requirements